跳转至正文
产品概述

概述

The Keithley 2601B-PULSE combines a high-current pulser with a conventional SourceMeter source measure unit for pulsed I-V characterization. The pulser is designed for testing devices such as VCSELs, laser diodes, LEDs, and other semiconductor components. Its pulse output and ordinary SMU operation have separate electrical and measurement limits.

The pulser supplies bipolar current pulses up to 10 A at 10 V, with programmable pulse widths from 10 to 500 µs. At 10 A, the maximum duty cycle is 3% when the bias-current magnitude is 10 mA or less. Pulse timing, bias current, cable and DUT inductance, and the published thermal limits must be considered together. The specified total cable-plus-DUT inductance limit is 3 µH, measured at 100 kHz.

Two 1 MS/s, 18-bit digitizers capture pulse voltage and current in the pulser function. Effective resolution depends on the measurement aperture; this is not a claim of 18 effective bits at 1 MS/s. Ordinary SMU operation uses its separate measurement specifications and provides nominal DC operating points of 40 V at 1 A or 6 V at 3 A, with a published maximum DC power of 40.4 W. Full-power sourcing is specified through 28 °C ambient, with additional derating above 28 °C or in sink operation.

Embedded TSP test scripts, TSP-Link, GPIB, rear USB device control, RS-232, Ethernet, and digital I/O support automated test integration. The front USB host port serves mass-storage devices. Pulser accuracy specifications require remote sensing and the other stated manufacturer conditions.

Manufacturer image reference: The photograph shows the exact 2601B-PULSE model. It does not establish that an external interlock/connector box, test fixture, cables, or other accessories are included with the offered unit.

技术信息

技术规格

Pulser Output and Timing

Maximum pulse operating point
Bipolar current pulses up to ±10 A at ±10 V; separate from continuous SMU DC limits
10 A pulse duty cycle
Maximum 3% with bias-current magnitude ≤10 mA; full-power sourcing through 28 °C ambient, with thermal derating above 28 °C or in sink operation
Programmable pulse width
10 µs to 500 µs
Pulse-width programming resolution
1 µs
Minimum programmed off-time
16 µs; duty-cycle limits also apply
Maximum load inductance
3 µH total cable plus DUT inductance, measured at 100 kHz
Bias-current source range
250 mA; achievable pulse duty cycle depends on bias current

Pulser and SMU Measurement

High-speed pulser digitizers
Two 1 MS/s, 18-bit digitizers for pulse voltage and current; effective bits depend on aperture and are not 18 bits at 1 MS/s
Pulser voltage measurement ranges
5 V and 10 V
Pulser current measurement ranges
1 A, 5 A, and 10 A
SMU voltage ranges
100 mV, 1 V, 6 V, and 40 V
SMU current ranges
100 nA through 3 A; the 10 A range requires SMU extended range mode. The 100 nA range is unavailable in High Capacitance Mode
Sense configurations
SMU: remote sensing or properly zeroed local operation. Pulser accuracy specifications: four-wire remote sensing

DC Operation and Automation

SMU functions
Voltage or current source and measurement; four-quadrant source or sink operation
Nominal SMU DC operating points
40 V at 1 A or 6 V at 3 A; separate from the pulser output envelope
Maximum SMU DC output power
40.4 W; full-power sourcing through 28 °C ambient, with additional derating above 28 °C or in sink operation
Embedded test control
Test Script Processor (TSP) for stored instrument test sequences
System expansion
TSP-Link communication and triggering with compatible instruments
Host and storage interfaces
IEEE-488/GPIB, rear USB 2.0 device/TMC488, RS-232, and 10/100BaseT Ethernet; front USB 2.0 host for mass storage
Digital I/O
14 open-drain I/O bits on a 25-pin D connector

销售与服务

保修
6个月保修
校准状态
可提供校准证书或测试报告(如报价单中所述)
制造商列出的标准配件
2601B-P-INT Rear panel interlock and cable connector box; 7709-308A Digital I/O connector; 2 × 17469460X TSP-Link/Ethernet cable; CS-1616-3 Safety interlock mating connector; 2601B-PULSE Quick Start Guide.

为什么选择与我们合作?

我们会与您核对相关细节

从规格到配件,我们帮助您了解产品包含哪些内容。.

清楚自己买的是什么

我们会分享我们对该病情的了解,以及所有可提供的记录。.

请告诉我们您的需求

有具体的要求或截止日期吗?我们将专注于您关心的重点。.

没必要从头开始

我们会记录您的提问和之前的讨论内容,因此您可以轻松地从上次讨论的地方继续。.

A closer look at 2601B-PULSE

A closer look at 2601B-PULSE
咨询

咨询此产品

请告知您的需求,我们将根据当前的产品背景进行评估。.

请输入电子邮箱地址或电话号码。.

请使用您的企业邮箱,然后输入我们发送到该邮箱的6位验证码。.