The Keithley 2601B-PULSE combines a high-current pulser with a conventional SourceMeter source measure unit for pulsed I-V characterization. The pulser is designed for testing devices such as VCSELs, laser diodes, LEDs, and other semiconductor components. Its pulse output and ordinary SMU operation have separate electrical and measurement limits.
The pulser supplies bipolar current pulses up to 10 A at 10 V, with programmable pulse widths from 10 to 500 µs. At 10 A, the maximum duty cycle is 3% when the bias-current magnitude is 10 mA or less. Pulse timing, bias current, cable and DUT inductance, and the published thermal limits must be considered together. The specified total cable-plus-DUT inductance limit is 3 µH, measured at 100 kHz.
Two 1 MS/s, 18-bit digitizers capture pulse voltage and current in the pulser function. Effective resolution depends on the measurement aperture; this is not a claim of 18 effective bits at 1 MS/s. Ordinary SMU operation uses its separate measurement specifications and provides nominal DC operating points of 40 V at 1 A or 6 V at 3 A, with a published maximum DC power of 40.4 W. Full-power sourcing is specified through 28 °C ambient, with additional derating above 28 °C or in sink operation.
Embedded TSP test scripts, TSP-Link, GPIB, rear USB device control, RS-232, Ethernet, and digital I/O support automated test integration. The front USB host port serves mass-storage devices. Pulser accuracy specifications require remote sensing and the other stated manufacturer conditions.
Manufacturer image reference: The photograph shows the exact 2601B-PULSE model. It does not establish that an external interlock/connector box, test fixture, cables, or other accessories are included with the offered unit.
사양
Pulser Output and Timing
- Maximum pulse operating point
- Bipolar current pulses up to ±10 A at ±10 V; separate from continuous SMU DC limits
- 10 A pulse duty cycle
- Maximum 3% with bias-current magnitude ≤10 mA; full-power sourcing through 28 °C ambient, with thermal derating above 28 °C or in sink operation
- Programmable pulse width
- 10 µs to 500 µs
- Pulse-width programming resolution
- 1 µs
- Minimum programmed off-time
- 16 µs; duty-cycle limits also apply
- Maximum load inductance
- 3 µH total cable plus DUT inductance, measured at 100 kHz
- Bias-current source range
- 250 mA; achievable pulse duty cycle depends on bias current
Pulser and SMU Measurement
- High-speed pulser digitizers
- Two 1 MS/s, 18-bit digitizers for pulse voltage and current; effective bits depend on aperture and are not 18 bits at 1 MS/s
- Pulser voltage measurement ranges
- 5 V and 10 V
- Pulser current measurement ranges
- 1 A, 5 A, and 10 A
- SMU voltage ranges
- 100 mV, 1 V, 6 V, and 40 V
- SMU current ranges
- 100 nA through 3 A; the 10 A range requires SMU extended range mode. The 100 nA range is unavailable in High Capacitance Mode
- Sense configurations
- SMU: remote sensing or properly zeroed local operation. Pulser accuracy specifications: four-wire remote sensing
DC Operation and Automation
- SMU functions
- Voltage or current source and measurement; four-quadrant source or sink operation
- Nominal SMU DC operating points
- 40 V at 1 A or 6 V at 3 A; separate from the pulser output envelope
- Maximum SMU DC output power
- 40.4 W; full-power sourcing through 28 °C ambient, with additional derating above 28 °C or in sink operation
- Embedded test control
- Test Script Processor (TSP) for stored instrument test sequences
- System expansion
- TSP-Link communication and triggering with compatible instruments
- Host and storage interfaces
- IEEE-488/GPIB, rear USB 2.0 device/TMC488, RS-232, and 10/100BaseT Ethernet; front USB 2.0 host for mass storage
- Digital I/O
- 14 open-drain I/O bits on a 25-pin D connector
판매 및 서비스
- 보증
- 6개월 보증
- 보정 상태
- 교정 증명서 또는 시험 보고서 제공 가능 (견적서에 명시된 바와 같이)
- 제조사에서 명시한 표준 액세서리
- 2601B-P-INT Rear panel interlock and cable connector box; 7709-308A Digital I/O connector; 2 × 17469460X TSP-Link/Ethernet cable; CS-1616-3 Safety interlock mating connector; 2601B-PULSE Quick Start Guide.
왜 저희와 함께 일해야 할까요?
고객님과 함께 세부 사항을 확인합니다
사양부터 액세서리까지, 어떤 구성품이 포함되어 있는지 자세히 안내해 드립니다.
무엇을 구매하는지 정확히 파악하세요
저희는 해당 질환에 대해 알고 있는 정보와 확보된 기록을 모두 공유해 드립니다.
필요한 사항을 알려주세요
특별한 요구 사항이나 마감일이 있으신가요? 고객님께 중요한 사항에 집중해 드리겠습니다.
다시 시작할 필요는 없습니다
저희는 여러분의 질문과 이전 토론 내용을 기록해 두므로, 중단했던 부분부터 다시 이어가기 쉽습니다.
A closer look at 2601B-PULSE

이 제품에 대해 문의하기
요구 사항을 알려주시면 현재 제품 상황을 검토해 드리겠습니다.
감사합니다
귀하의 문의 내용이 등록되었습니다. 여기에 표시된 확인 번호를 잘 보관해 주십시오.