跳转至正文
产品概述

概述

The Keithley 2400 is a single-channel SourceMeter source measure unit for DC component characterization. It combines a programmable voltage or current source with voltage, current, and resistance measurements in one instrument. Four-quadrant operation supports both sourcing and sinking within the model’s operating limits.

The 2400 is nominally rated at 20 W, with operating points of 200 V at 100 mA or 20 V at 1 A. Voltage and current maxima apply to different parts of the operating envelope. Continuous high-current sourcing, sink operation, and elevated ambient temperature require the derating described in the manufacturer documentation.

Local and remote sensing, guarded resistance measurements, and linear, logarithmic, or custom sweeps support leakage, diode, LED, resistor, and other I-V tests. The source-memory sequencer stores up to 100 instrument configurations, while reading buffers hold up to 5,000 readings at five digits. GPIB, RS-232, Trigger Link, and digital I/O support integration with automated test systems.

The data sheet identifies the 2400 as a legacy model. The ranges below come from the actual 2400 rows; higher-voltage and higher-current ratings elsewhere in the Series 2400 document belong to other models. Resolution values are range-dependent and do not state measurement accuracy.

Manufacturer image reference: The photograph shows a Keithley 2400 as a model reference. It does not establish the offered unit’s appearance or included accessories.

技术信息

技术规格

DC Source and Operating Envelope

SMU channels
1
Nominal DC power rating
20 W; continuous sink and high-current operation are subject to the manufacturer thermal and duty-cycle limits
Nominal voltage capability
200 V; available current depends on the operating point
Nominal current capability
1 A; available voltage depends on the operating point
Nominal operating points
200 V at 100 mA or 20 V at 1 A, with either polarity
Source and sink operation
Four quadrants; source voltage or current, with programmed compliance limits

Measurement and Sensing

Measurement functions
Voltage, current, and resistance
Measurement resolution
Up to 6.5 digits; selected speed and range affect resolution
Voltage ranges
200 mV, 2 V, 20 V, and 200 V
Current ranges
1 µA, 10 µA, 100 µA, 1 mA, 10 mA, 100 mA, and 1 A
Lowest-range voltage measurement resolution
1 µV on the 200 mV range; resolution is not accuracy
Lowest-range current measurement resolution
10 pA on the 1 µA range; resolution is not accuracy

Test Sequencing and Interfaces

Remote interfaces
IEEE-488/GPIB (SCPI 1995.0), RS-232, and Keithley Trigger Link
Sweep modes
Linear staircase, logarithmic staircase, and custom sweeps
Source-memory sequence
Up to 100 instrument configurations
Reading memory
5,000 readings at five digits; two 2,500-point buffers
Digital I/O
One trigger input and four TTL/relay-drive outputs
Sense connections
Local two-wire, remote four-wire, and guarded six-wire resistance configurations

销售与服务

保修
6个月保修
校准状态
可提供校准证书或测试报告(如报价单中所述)
制造商列出的标准配件
Test leads.

为什么选择与我们合作?