The Keithley 2400 is a single-channel SourceMeter source measure unit for DC component characterization. It combines a programmable voltage or current source with voltage, current, and resistance measurements in one instrument. Four-quadrant operation supports both sourcing and sinking within the model’s operating limits.

The 2400 is nominally rated at 20 W, with operating points of 200 V at 100 mA or 20 V at 1 A. Voltage and current maxima apply to different parts of the operating envelope. Continuous high-current sourcing, sink operation, and elevated ambient temperature require the derating described in the manufacturer documentation.

Local and remote sensing, guarded resistance measurements, and linear, logarithmic, or custom sweeps support leakage, diode, LED, resistor, and other I-V tests. The source-memory sequencer stores up to 100 instrument configurations, while reading buffers hold up to 5,000 readings at five digits. GPIB, RS-232, Trigger Link, and digital I/O support integration with automated test systems.

The data sheet identifies the 2400 as a legacy model. The ranges below come from the actual 2400 rows; higher-voltage and higher-current ratings elsewhere in the Series 2400 document belong to other models. Resolution values are range-dependent and do not state measurement accuracy.

Manufacturer image reference: The photograph shows a Keithley 2400 as a model reference. It does not establish the offered unit’s appearance or included accessories.

Technische Informationen

Technische Daten

DC Source and Operating Envelope

SMU channels
1
Nominal DC power rating
20 W; continuous sink and high-current operation are subject to the manufacturer thermal and duty-cycle limits
Nominal voltage capability
200 V; available current depends on the operating point
Nominal current capability
1 A; available voltage depends on the operating point
Nominal operating points
200 V at 100 mA or 20 V at 1 A, with either polarity
Source and sink operation
Four quadrants; source voltage or current, with programmed compliance limits

Measurement and Sensing

Measurement functions
Voltage, current, and resistance
Measurement resolution
Up to 6.5 digits; selected speed and range affect resolution
Voltage ranges
200 mV, 2 V, 20 V, and 200 V
Current ranges
1 µA, 10 µA, 100 µA, 1 mA, 10 mA, 100 mA, and 1 A
Lowest-range voltage measurement resolution
1 µV on the 200 mV range; resolution is not accuracy
Lowest-range current measurement resolution
10 pA on the 1 µA range; resolution is not accuracy

Test Sequencing and Interfaces

Remote interfaces
IEEE-488/GPIB (SCPI 1995.0), RS-232, and Keithley Trigger Link
Sweep modes
Linear staircase, logarithmic staircase, and custom sweeps
Source-memory sequence
Up to 100 instrument configurations
Reading memory
5,000 readings at five digits; two 2,500-point buffers
Digital I/O
One trigger input and four TTL/relay-drive outputs
Sense connections
Local two-wire, remote four-wire, and guarded six-wire resistance configurations

Vertrieb und Service

Garantie
6 Monate Garantie
Kalibrierungsstatus
Kalibrierzertifikat oder Prüfbericht verfügbar (wie in Ihrem Angebot angegeben)
Vom Hersteller angegebenes Standardzubehör
Test leads.

Warum sollten Sie bei uns arbeiten?

Wir klären die Einzelheiten mit Ihnen ab

Von den technischen Daten bis hin zum Zubehör – wir helfen Ihnen dabei, den Überblick zu behalten, was alles im Lieferumfang enthalten ist.

Wissen Sie, was Sie kaufen

Wir geben unser Wissen über die Erkrankung weiter und stellen alle verfügbaren Unterlagen zur Verfügung.

Sagen Sie uns, was Sie brauchen

Haben Sie spezielle Anforderungen oder einen bestimmten Termin? Wir konzentrieren uns auf das, was Ihnen wichtig ist.

Man muss nicht von vorne anfangen

Wir speichern Ihre Fragen und bisherigen Diskussionen, sodass wir problemlos dort weitermachen können, wo wir aufgehört haben.

A closer look at 2400

A closer look at 2400
Anfrage

Fragen Sie nach diesem Produkt

Teilen Sie uns Ihre Anforderungen mit, und wir werden den aktuellen Produktkontext prüfen.

Geben Sie eine E-Mail-Adresse oder eine Telefonnummer ein.

Geben Sie Ihre geschäftliche E-Mail-Adresse ein und anschließend den 6-stelligen Code, den wir Ihnen dorthin senden.