Übersicht
The Keithley 2612B is a dual-channel System SourceMeter source measure unit with isolated channels for I-V characterization and automated component testing. Each channel combines voltage or current sourcing with electrical measurement and four-quadrant source or sink operation.
The published DC envelope includes 200 V at 100 mA and 20 V at 1.5 A per channel, with a maximum DC power specification of 30.3 W per channel. A separate pulse region permits 10 A at 5 V for up to 1 ms at a maximum 2.2% duty cycle. These voltage and current limits are operating points rather than simultaneous maxima. Full-power sourcing is specified through 30 °C ambient; higher ambient temperatures and sink operation require additional derating.
Embedded Test Script Processor control runs stored test sequences, and TSP-Link supports coordination with compatible instruments. Contact Check, digital I/O, GPIB, rear USB device control, RS-232, and Ethernet support system integration. The front USB host connection serves mass-storage devices. Local or four-wire remote sensing can be selected to suit the test setup.
The specification rows below use the 2611B/2612B/2614B section of the official 2600B data sheet, with the 2612B-specific channel and feature distinctions retained. The 100 fA default current display resolution applies to the 100 nA range in single-channel display mode; that range is unavailable in High Capacitance Mode. Resolution does not state measurement accuracy.
Manufacturer image reference: The official group photograph shows 2611B, 2612B, and 2614B instruments together. This listing is for the Keithley 2612B; the photograph does not represent a multi-instrument package or establish the offered unit’s accessories.
Technische Daten
Channels and Source Envelope
- SMU channels
- 2 isolated channels
- Source and sink operation
- Four-quadrant voltage or current sourcing and measurement
- Nominal DC operating points per channel
- 200 V at 100 mA or 20 V at 1.5 A; voltage and current maxima are not simultaneous
- Maximum DC output power
- 30.3 W per channel; full-power sourcing through 30 °C ambient, with additional derating above 30 °C or in sink operation
- 10 A pulse operating point
- 10 A at 5 V; maximum 1 ms pulse width and 2.2% duty cycle. Thermal limits apply in sink mode and above 30 °C ambient
- Lowest-range programming resolution
- 5 µV on the 200 mV voltage range; 2 pA on the 100 nA current range
Measurement and Sense Connections
- Measurement functions
- Voltage and current
- Voltage measurement ranges
- 200 mV, 2 V, 20 V, and 200 V
- Current measurement ranges
- 100 nA through 1.5 A; 10 A range is available only in pulse mode. The 100 nA range is unavailable in High Capacitance Mode
- Lowest-range voltage display resolution
- 100 nV on the 200 mV range in single-channel display mode; resolution is not accuracy
- Lowest-range current display resolution
- 100 fA on the 100 nA range in single-channel display mode; resolution is not accuracy
- Sense connections
- Local or four-wire remote sensing
Automation and System Interfaces
- Embedded test control
- Test Script Processor (TSP) for stored instrument test sequences
- System expansion
- TSP-Link triggering and communication with compatible instruments
- Host and storage interfaces
- IEEE-488/GPIB, rear USB 2.0 device/TMC488, RS-232, and 10/100 Ethernet; front USB 2.0 host for mass storage
- Contact Check
- Measures SENSE HI-to-HI and SENSE LO-to-LO contact resistances
- Digital I/O
- 14 open-drain I/O bits on a 25-pin D connector
- 200 V safety interlock
- External safety-interlock input; 200 V operation depends on the interlock state
Vertrieb und Service
- Garantie
- 6 Monate Garantie
- Kalibrierungsstatus
- Kalibrierzertifikat oder Prüfbericht verfügbar (wie in Ihrem Angebot angegeben)
- Vom Hersteller angegebenes Standardzubehör
- 2 × 2600-KIT screw terminal connector, strain relief, and cover for one SMU channel; 7709-308A Digital I/O connector; 2 × 17469460X TSP-Link/Ethernet cable; Operator and programming manuals.
Warum sollten Sie bei uns arbeiten?
Wir klären die Einzelheiten mit Ihnen ab
Von den technischen Daten bis hin zum Zubehör – wir helfen Ihnen dabei, den Überblick zu behalten, was alles im Lieferumfang enthalten ist.
Wissen Sie, was Sie kaufen
Wir geben unser Wissen über die Erkrankung weiter und stellen alle verfügbaren Unterlagen zur Verfügung.
Sagen Sie uns, was Sie brauchen
Haben Sie spezielle Anforderungen oder einen bestimmten Termin? Wir konzentrieren uns auf das, was Ihnen wichtig ist.
Man muss nicht von vorne anfangen
Wir speichern Ihre Fragen und bisherigen Diskussionen, sodass wir problemlos dort weitermachen können, wo wir aufgehört haben.