AQ6374 Overview and Applications

The Yokogawa AQ6374 measures spectra across visible and near-infrared wavelengths. It is suited to laser and broadband-source characterization and wavelength-dependent fiber-loss measurements. Eight resolution settings let engineers balance spectral detail and measurement needs.

Measurement and Analysis

Seven traces can be displayed together for comparison. Built-in analysis covers laser, LED, optical-filter, and color measurements. A purge connection helps reduce water-vapor absorption within the monochromator. The internal reference source supports optical alignment and wavelength calibration.

Applications and Automation

  • Characterize visible-to-telecom laser sources and broadband emitters.
  • Measure optical-fiber loss versus wavelength with a suitable external source.
  • Evaluate visible LED spectra and color.
  • Integrate remote measurements using GP-IB, RS-232, or Ethernet; Ethernet supports TCP/IP and VXI-11 or socket control.

Model and Configuration Notes

Yokogawa lists the AQ6374 as discontinued in November 2023 and identifies the AQ6374E as its successor. This listing describes the AQ6374. Confirm the unit’s connector adapters, supplied accessories, condition, and calibration records with your quotation.

Specification conditions: [A] 0.05 nm resolution. [A–C] 9.5/125 µm SM fiber after internal alignment, without purge gas. [B] Absolute level display, resolution ≥0.2 nm. [C] Specified PC-polished SM fiber. Sensitivity and sweep figures assume pulse mode off. [D] 1,001 points, one average; excludes 570–580 and 900–1000 nm. Performance temperature: 18–28°C. Follow the English guide for complete conditions, warm-up, and calibration.

技術情報

AQ6374 Technical Specifications

Optical Performance

Wavelength range
350–1750 nm
Resolution settings
0.05, 0.1, 0.2, 0.5, 1, 2, 5, and 10 nm
Wavelength accuracy
±0.2 nm overall; ±0.05 nm at 1523 nm and at 633 nm after 633 nm HeNe calibration [A]
Wavelength repeatability
±0.015 nm over 1 minute [A]
Minimum wavelength sampling interval
0.002 nm
Sampling points
101–100,001; AUTO available
Level sensitivity
−70 dBm at 400–900 nm; −80 dBm at 900–1600 nm; HIGH3 [B]
Maximum safe total input
+10 dBm at 400–550 nm; +20 dBm at 550–1750 nm [B]
Level accuracy
±1.0 dB at 1550 nm, −20 dBm input, HIGH1–HIGH3 [B,C]
Level linearity
±0.2 dB for −40 to 0 dBm input, HIGH1–HIGH3 [B]
Close-in dynamic range
60 dB at peak ±1 nm (633/1523 nm); 0.05 nm resolution [A]
Sweep time
0.5 s at NORM_AUTO, span ≤100 nm [D]

Fiber Input and Bench Integration

Supported fibers
SM 9.5/125 µm; GI 50/125 and 62.5/125 µm; large-core fibers up to 800 µm
Remote interfaces
GP-IB, RS-232, Ethernet (TCP/IP); socket and VXI-11 Ethernet modes
Bench size and mass
Approx. 426 × 221 × 459 mm (W × H × D), excluding protectors/handle; approx. 19 kg

販売・サービス

保証
6ヶ月保証
校正状況
校正証明書または試験報告書をご用意しております(お見積書に記載の通り)
Manufacturer-Listed Standard Accessories
Original manufacturer list: one regional power cord (none with -Y); two A9088ZM rubber feet supplied as one sheet; one A1023US manual CD; one each of the Getting Started Guide, China document, safety manual and worldwide contact list. Manufacturer-listed standard accessories are a reference to the original model documentation. They do not guarantee the contents of a used instrument shipment. Confirm the actual supplied items for the specific unit before ordering.
ダウンロード

文書

なぜ当社と提携すべきなのでしょうか?

お客様と詳細を確認させていただきます

仕様から付属品に至るまで、何が含まれているのかをご理解いただけるようお手伝いいたします。.

何を買うのかをしっかり把握しましょう

その病状について私たちが把握している情報と、入手可能な記録をすべてお伝えします。.

ご要望をお聞かせください

具体的なご要望や納期はおありですか?お客様にとって重要な点に重点を置いて対応いたします。.

一からやり直す必要はありません

お客様の質問や過去のやり取りはすべて記録していますので、前回の続きからスムーズに再開できます。.

A closer look at AQ6374

A closer look at AQ6374
お問い合わせ

この商品についてお問い合わせください

ご要望をお聞かせください。弊社にて現在の製品状況を精査いたします。.

メールアドレスまたは電話番号を入力してください。.

会社のメールアドレスを入力し、そのメールアドレス宛に送信される6桁のコードを入力してください。.